Chapter 5 - Components and Functions (Theory)
5 - 45
5-11-2
Relay Board, RLY
5-11-2-1
General Description
The Relay Board has three PA probe connectors, a “blind” PA slot used to park a fourth probe and a
connector for a stand alone Doppler probe containing two transducer elements.
•
In receiving mode, the Relay Board connects the transducer channels from the selected probe to
the RX-128 and TX-128 boards via the XDBUS.
•
The Relay Board routes the TX power supplies to the active probe.
•
It also routes the Probe Multiplexer Control Signal to the active probe.
5-11-2-2
Location in the Unit
Figure 5-21 RLY Board: Location in Unit
RLY Board
Front of
Scanner