Chapter 4 - Functional Checks
4 - 3
4-2-1-2
Turn Unit ON
Table 4-3
Turn Unit ON
Step
Task
Expected Result(s)
1
Switch ON the Mains Power Circuit Breaker at the rear of the
unit.
You will hear a “click” from the relays in the AC Power/AC Controller. The
unit is ready for Boot Up.
2
Press the
ON/OFF
key on the Control Panel (Keyboard) once.
.
1) The unit’s ventilation fan starts on full speed, but slows down after a
few seconds (listen to the fan noise).
2) Power is distributed to the peripherals, Control Panel (Console),
Monitor, Card Rack and Backend Processor.
3) Backend Processor and rest of Scanner starts with the following
sequence:
4) Backend Processor is turned ON and starts to load the software.
5) The Start Screen (Vivid) is displayed on the monitor
6) A start-up bar is displayed on the monitor, indicating the progress of
the software loading.
NOTE:
If the unit has been in the OFF condition
for an extended period of time, (3 to 5
days or more), the unit may not boot, or
may beep when turned on.
See “System
Doesn’t Boot” on page 7-67.
for more
details, including how to recover.
Mains Power Circuit Breaker in ON position
ON/OFF
key