Section 4-3 - Functional Checks
4-3-4-4
Check Up/Down, Left/Right, B Color Maps and Cineloop
These 2D Mode Functions are located on Mode Dependant Keys (Push Buttons)
The results of these adjustments must be verified on the 2D sector on the screen.
Table 4-7
2D Mode: Check Up/Down, Left/Right, B Color Maps and Cineloop
Step
Task
Expected Result
1.
Press once on the assignable key labeled
UP/DOWN
Enables the 2D image to be flipped 180 degrees.
2.
Press once on the assignable key labeled
LEFT/RIGHT
Enables a mirror image of the 2D image to be created.
The left/right reference marker
V
moves to the other
side of the image.
3.
Select
B COLOR MAPS
Displays a 2D Map Menu on the screen to optimize the
greyscale presentation.
The menu enables an option from a list of non-linear
gray-curves or different 2D-colorized curves to be
selected.
4.
Press
FREEZE
Select
CINELOOP
)
Starts cine loop acquisition.