Chapter 5 - Components and Functions (Theory)
5 - 61
5-11-7
Transducer Bus Boards, XD BUS Boards
5-11-7-1
General Description
The two XD BUS Boards plugs into the rear edge connectors on the Relay Board, the TX128 Boars and
on the RX-128 Board.
The XD signals to and from the probes (via the Relay Board) are routed via this board.
5-11-7-2
Location in the Unit
5-11-7-3
Inputs
Input Signals
Figure 5-31 XD BUS Boards: Location in Unit
Table 5-44 Input Signals
Input
Description
Connection from:
XD1 - XD128
Back scattered echo signals (Receive Mode) and
Transmitter pulses (Transmit Mode)
Active Probe via Relay board in Receive
Mode
TX128 in Transmit Mode.
XD BUS Board
Front of
Scanner
XD BUS Board