Section 7-4 - Acquisition Diagnostics
7-4-4-2
System Test Screen Description
7-4-4-2-1
Test Loops, System
Figure 7-58 System Test Screen, Upper Part
Table 7-4
Test loops, System - Test All, System - Break
Field
Key
Description
Test loops
-
In some cases you may want to do the same test several times.
Enter a number to indicate how many times you want the test to repeat itself. The default
number is 1 (one).
System
Test All
Use this function to start a sequence of all the tests available in the Test Screen.
The diagnostic test is done in this sequence;
1.) PC2IP
2.) IMP (Image Port Board)
3.) VME (VME data bus on the Back Plane)
4.) SDP (Spectrum Doppler Processor)
5.) RFT (Radio Frequency & Tissue Board)
6.) FEC (Front End Controller Board)
7.) BF (Beam Former Boards)
8.) RX (Receiver Board)
9.) TXP (TX Power)
NOTE:
Relay Board (RLB) and Chain are not tested in this
test. The Transmitter Board (TX) don’t have its own
test, but some parts of the TX board is tested
during the FEC test.
NOTE:
If the system must be rebooted, do so. Continue
with the testing from the point at which you were
required to stop.
System Break
If you have started a looped test, you may click BREAK to stop the test.
The test will continue until the current test has finished.
System
Test Loops
Mode