GE M
EDICAL
S
YSTEMS
D
IRECTION
FC091194, R
EVISION
02
V
IVID
7 / V
IVID
7 PRO S
ERVICE
M
ANUAL
8 - 12
Section 8-3 - SW Loading Procedure
8-3-5
Prepare MO Disks for Image Storage
1.) Label the MO Disk sequentially.i.e Image 1A (Side A), Image 1B (side B), Image 2A etc.
NOTE:
When formatting the MO disk, give it a logical name and make sure to physically label the disk.
A labelling convention should be followed so that each MO disk gets a unique label. The
maximum numbers of label characters allowed is 11.
MO Disk Label side B must differ from side A.
2.) If the MO Disks is already formatted and labelled move to
section 8-3-6 on page 8-13
.
3.) Insert a MO disk, side A, into the empty MO drive.
4.) Press
CONFIG
(F2) on the alphanumeric Keyboard on the operator panel.
5.) From the Operator Login window select
adm
from the Operator pull-down menu.
6.) Select LOG ON.
7.) Select CONNECTIVITY in the Image and Analysis menu.
8.) Select TOOLS in the Connectivity menu.
9.) In the Label window, type the same name as labelled on the MO disk side A.
10.)Select FORMAT.
11.)A warning “Are you sure you want to format H: and remove all data from the media” appears, select
OK.
12.)Wait until the information dialogue window appears on screen.
13.)Select OK to continue.
14.)Verify the disk formatting status has been updated.
15.)Press
ALT-E.
16.)Select the
5 1/4 MO Disk
, to eject Disk.
17.)Insert the MO disk, side B and Repeat step 9 to 16.
Figure 8-1 Prepare MO Disk - Image Storage