Chapter 2 - Pre-Installation
2 - 3
2-2-2
Electrical Requirements
NOTE:
GE Medical Systems requires a dedicated power and ground for the proper operation of its Ultrasound
equipment. This dedicated power shall originate at the last distribution panel before the system.
Sites with a mains power system with defined Neutral and Live:
The dedicated line shall consist of one phase, a neutral (not shared with any other circuit), and a full
size ground wire from the distribution panel to the Ultrasound outlet.
Sites with a mains power system without a defined Neutral:
The dedicated line shall consist of one phase (two lines), not shared with any other circuit, and a full
size ground wire from the distribution panel to the Ultrasound outlet.
Please note that image artifacts can occur, if at any time within the facility, the ground from the main
facility's incoming power source to the Ultrasound unit is only a conduit.
2-2-2-1
Electrical Requirements for Vivid 7 and Vivid 7 PRO
Electrical Specifications for Vivid 7 and Vivid 7 PRO. Monitor and on board peripherals are included.
2-2-2-2
Inrush Current
•
230 VAC: 12 A
•
120 VAC: 7 A
2-2-2-3
Site Circuit Breaker
It is recommended that the branch circuit breaker for the unit be readily accessible.
2-2-2-4
Site Power Outlets
A dedicated AC power outlet must be within reach of the unit without extension cords. Other outlets
adequate for the external peripherals, medical and test equipment needed to support this unit must also
be present within 1 m (3.2 ft.) of the unit. Electrical installation must meet all current local, state, and
national electrical codes.
2-2-2-5
Unit Power Plug
If the unit arrives without a power plug, or with the wrong plug, you must contact your GE dealer or the
installation engineer must supply what is locally required.
Table 2-4
Electrical Specifications for Vivid 7 / Vivid 7 PRO
GE VINGMED
PART NUMBER
DESCRIPTION
VOLTAGE
TOLERANCES
CURRENT
FREQUENCY
FB000030
VIVID 7 (BT ’01)
230 VAC
±10%
5 A
50-60 Hz
FC000060
VIVID 7 (BT ’01)
115 VAC
±10%
10 A
50-60 Hz
FC000180
VIVID 7 PRO (BT ’02)
230 VAC
±10%
5 A
50-60 Hz
FC000190
VIVID 7 PRO (BT ’02)
115 VAC
±10%
10 A
50-60 Hz
FC000200
VIVID 7 (BT ’02)
230 VAC
±10%
5 A
50-60 Hz
FC000210
VIVID 7 (BT ’02)
115 VAC
±10%
10 A
50-60 Hz
CAUTION
POWER OUTAGE MAY OCCUR. The Vivid 7 / Vivid 7 PRO requires a dedicated single branch
circuit. To avoid circuit overload and possible loss of critical care equipment, make sure you
DO NOT have any other equipment operating on the same circuit.