Chapter 8 - Replacement Procedures
8 - 11
8-3-4
Preparation
Table 8-5
Turn the System ON
Step
Task
Expected Result(s)
1.
Switch ON the Mains Power Circuit Breaker at the rear of the
System.
You will hear a “click” from the relays in the AC Power/AC Controller. The
system is ready for Boot Up.
2.
Press
ON/OFF
key on the Control Panel (Keyboard) once.
.
1). System’s ventilation fan starts on full speed, but slows down after a
few seconds (listen to the fan noise).
2). Power is distributed to the peripherals, Control Panel, Monitor, Card
Rack and Backend Processor.
3). Backend Processor and rest of Scanner starts with the following
sequence:
4). Backend Processor is turned ON and starts to load the software.
5). The Start Screen (Vivid) is displayed on the monitor
6). A start-up bar is displayed on the monitor, indicating the progress of
the software loading.
Mains Power Circuit Breaker on
ON/OFF