GE M
EDICAL
S
YSTEMS
D
IRECTION
FC091194, R
EVISION
02
V
IVID
7 / V
IVID
7 PRO S
ERVICE
M
ANUAL
Chapter 4 - Functional Checks
4 - 55
4-3-15-4
Check First, Last Cycle, Sync and Select all
4-3-15-5
Adjust Scroll and Cine Speed
Table 4-34 Cineloop Controls Checks
Step
Task
Expected Result(s)
1.
Press
FIRST
and
LAST CYCLE
Select either the first or the last heart cycle.
2.
Press
SYNC
Phase synchronize multiple loops.
3.
Press
SELECT ALL
Moves the cineloop boundaries to include all heart
cycles.
Table 4-35 Cineloop Controls Checks
Step
Task
Expected Result(s)
1.
Adjust
Scroll
(in Frozen mode):
Press
TRACKBALL
until the Scroll trackball assignable
is selected.
Use the Trackball.
Move the current frame marker and review images.
2.
Adjust Cine Speed (in Cine Replay mode):
Press
FREEZE
to freeze the image.
Press
CINELOOP
.
Roll the Trackball to the left.
Roll the Trackball to the right.
Step : the cine speed is decreased.