Section 4-3 - Functional Checks
4-3-15-3
Check Left Marker, Right Marker, Cycle Select and Number of Cycles
Table 4-33 Cineloop Controls Checks
Step
Task
Expected Result(s)
1.
Adjust
LEFT MARKER
and
RIGHT MARKER
Moves the Left and Right Markers to expand or trim the
cineloop boundaries.
2.
Adjust
CYCLE SELECT
Moves the cineloop boundaries from heart cycle to
heart cycle.
3.
Adjust
NUMBER OF CYCLE
Increase or decrease the number of heart cycles to be
included in the cineloop.