GE M
EDICAL
S
YSTEMS
D
IRECTION
FC091194, R
EVISION
02
V
IVID
7 / V
IVID
7 PRO S
ERVICE
M
ANUAL
8 - 8
Section 8-2 - Plastic Parts Replacement Procedures
8-2-6
Lower Rear Cover Replacements Procedure
8-2-6-1
Manpower
One person, 30 minutes
8-2-6-2
Tools
Phillips screwdriver size 2
8-2-6-3
Preparations
Shut down the system as described in
"Power Shut Down" on page 4-5
.
8-2-6-4
Lower Rear Cover Removal Procedure
1.) Remove Filter Cover shown in
"Filter Cover Removal Procedure" on page 8-6
.
2.) Remove both Side Covers as described in
"Side Covers Removal Procedure" on page 8-3
.
3.) Unscrew two (2) screws with magnetic lock (part number 080X1200) for Filter Cover. Left screw
contains also 3 or 4 washers (080C2053), right screw 5 or 6 washers, depending on thickness of
Lower Rear Cover.
4.) Unscrew four (4) remaining Phillips screws that fasten Lower Rear Cover to the chassis (See
Figure
8-4
).
5.) Remove Lower Rear Cover.
8-2-6-5
Lower Rear Cover Installation Procedure
1.) Align Lower Rear Cover with the chassis and fasten it with eight (8) Phillips screws.
2.) Install Filter Cover as described in
"Filter Cover Installation Procedure" on page 8-6
and Side
Covers as described in
"Side Covers Installation Procedure" on page 8-4
.
3.) New Lower Rear Cover: Install correct label, see
"Labels Locations" on page 1-14.
Figure 8-8 Lower Rear Cover Fastening Locations.
Screws
Lower Rear Cover
Screws
Magnetic lock
5 - 6 washers
3 - 4 washers
Содержание FC091194
Страница 2: ......
Страница 508: ...GE MEDICAL SYSTEMS DIRECTION FC091194 REVISION 02 VIVID 7 VIVID 7 PRO SERVICE MANUAL Index 6 Index ...
Страница 509: ......