GE M
EDICAL
S
YSTEMS
D
IRECTION
FC091194, R
EVISION
02
V
IVID
7 / V
IVID
7 PRO S
ERVICE
M
ANUAL
5 - 66
Section 5-11 - Circuit Boards
5-11-9
Spectrum Doppler Processor Board, SDP
5-11-9-1
General Description
•
This board performs Spectrum Analysis on the Doppler signals and makes the velocity traces,
•
The Doppler audio channels are generated on this board.
•
During Duplex modes (e.g. 2D and Doppler), The Missing Signal Estimator will fill the gaps in the
Doppler spectrum by stretching the “real” spectrum out to fill the missing time.
•
A serial link from the Internal I/O board, feeds the A/D converted traces (like ECG, Phono and
Pressure) into the SDP. These signals are sent together with the Doppler data via the Pipelink bus
to the Image Port Board.
5-11-9-2
Location in the Unit
Figure 5-34 SPD Board: Location in Unit
SDP Board
Front of
Scanner