Chapter 8 - Replacement Procedures
8 - 17
8-3-8 Backup of Patient Database and User Presets (cont’d)
.
NOTE:
No process indicator will appear on screen. Verify that the “Finished, OK” Status appears in the
Result column, and that the date appearing in the Last Successful Backup column is the current
date and time. The backup is completed.
6.) Remove the Backup MO Disk when the backup is completed.
Figure 8-13 Backup Warning