Section 8-4 - FRU Replacement Procedures
8-4-4
TX Power Replacement Procedure
8-4-4-1
Manpower:
One person, 30 minutes
8-4-4-2
Tools:
•
Phillips screwdriver 2
•
4 mm blade screwdriver
8-4-4-3
Preparations
1.) Shut down the system as described in
"Power Shut Down" on page 4-5
.
2.) Remove Right Side Cover as described in
"Side Covers Removal Procedure" on page 8-3
.
3.) Remove the Card Rack Door.
a.) Turn seven (7) quarter-turn screws to loosen Card Rack door.
b.) Holding the door, lift it out of its retainers along the lower edge of Card Rack.
4.) Disconnect PCI cable from IMP2 Board shown in
Figure 8-32
.
.
DANGER
Dangerous voltages are present in this equipment. Completely power off and
shut down the scanner before servicing.
Figure 8-32 DC Power, TX Power and Cabling.
HV power cable
DC power cable
PCI cable
IMP2 Board
TX Power
Retaining clasp
TX Power
Screws
DC Power