2 - 2
Section 2-2 - General Console Requirements
Section 2-2
General Console Requirements
2-2-1
Console Environmental Requirements
2-2-1-1
Cooling
The cooling requirement for the Vivid 7 / Vivid 7 PRO is 3500 BTU/hr. This figure does not include
cooling needed for lights, people, or other equipment in the room. Each person in the room places an
additional 300 BTU/hr. demand on the cooling system.
2-2-1-2
Lighting
Bright light is needed for system installation, updates and repairs. However, operator and patient
comfort may be optimized if the room light is subdued and indirect. Therefore a combination lighting
system (dim/bright) is recommended. Keep in mind that lighting controls and dimmers can be a source
of EMI which could degrade image quality. These controls should be selected to minimize possible
interference.
CAUTION
If the unit is very cold or hot, do not turn on its power until it has had a chance to acclimate to
its operating environment.
Table 2-2
Time for Settlement
°
C
60
55
50
45
40
35
30
25
20
15
10
5
0
-5
-10
-15
-20
-25
-30
-35
-40
°
F
140
131
122
113
104
96
86
77
68
59
50
41
32
23
14
5
-4
-13
-22
-31
-40
hrs
8
6
4
2
0
0
0
0
0
0
0
2
4
6
8
10
12
14
16
18
20
Table 2-3
Environmental Specifications for Vivid 7 / Vivid 7 PRO Scanners
Operating temperature
Storage temperature
Humidity
Heat dissipation
10 to 35
o
C (50 to 95
o
F)
-20 to 50
o
C (- 4 to 122
o
F)
< 90% rH non-condensing
3500 BTU pr hour