10 - Menu and dialog descriptions
327
Entry
Standard
Description
Parity
O
Defines the type of parity check of the serial interface between the
PC and the balance.
N= no parity check
E= even parity check
O= odd parity check.
The following table describes the entries under the "BalanceInit" section:
Entry
Standard
Description
StartString
----
Defines the character with which a weighing datum starts. This is
used for correct interpretation of the characters received.
EndString
----
Defines the character with which a weighing datum ends. This is
used for correct interpretation of the characters received.
WeightFactor
1
The WeightFactor allows correction of the weighing data sent by
the balance. The program expects weighing data milligram format.
For example, if the balance sends data in gram format, you must
set WeightFactor to 1000.
The following table describes the entries under the "PGP" section:
Entry
Standard
Description
ExecutablePath
---
Defines what signature program to use for signing and verifying
documents.
Sign Parameters
---
Defines what parameters must be sent to the signature program
when signing.
Verify Parameters ---
Defines what parameters must be sent to the signature program
when verifying.
The following table describes the entries under the "Menu" section:
Entry
Standard
Description
xxx
1,2,3
The text in this entry describes the respective menu command.
The value to the right of the equals sign defines the access level
with which the user may access this command. 1 means the
lowest, 3 is the highest level.
The following table describes the entries under the "System" section:
Entry
Standard
Description
Logon Timeout
60
Defines the time that must expire before the program displays the
login dialog again.
User1
---
Specifies the name of the user group which the instrument
software regards as level 1 users.
User2
Specifies the name of the user group which the instrument
software regards as level 2 users.
User3
Specifies the name of the user group which the instrument
software regards as level 3 users.
Further information
Further information includes instructions in which the dialog is used and knowledge articles
concerning the dialog:
Defining the analyzer / PC interface
(on page 84).
Creating new sections
(on page 77).
Granting authorizations
(on page 81).
Defining the autoexport directory
(on page 93).
Содержание DuMaster D-480
Страница 1: ...DuMaster D 480 Operation Manual 11593626C en...
Страница 4: ......
Страница 12: ......
Страница 16: ......
Страница 30: ......
Страница 102: ......
Страница 181: ...Operating instructions DuMaster D 480 B CHI Labortechnik AG 8 Maintaining the instrument 181...
Страница 286: ...Operating instructions DuMaster D 480 B CHI Labortechnik AG 10 Menu and dialog descriptions 286...
Страница 338: ......
Страница 352: ......
Страница 353: ......