Operating instructions DuMaster D-480
©BÜCHI Labortechnik AG
8 - Maintaining the instrument
193
Check the brass wool of the post combustion tube plug. If necessary, replace the brass wool.
Fill the Regainer by means a metal funnel into the post combustion tube (2).
To check the fill level use the rod, market with "full" and "empty".
Tap slightly against the post combustion tube after refilling the Regainer. Thus, the Regainer will
be compacted in the post combustion tube.
Phase 3: Insert of the post combustion tube
Top view furnace
The following picture shows a top view of the furnace pulled out:
Legend:
1 Post combustion tube plug
2 Post combustion tube
3 Combustion tube connection
4 Combustion tube
5 Plug of Reductor
6 Reductor
Push the post combustion tube plug (1) onto the post combustion tube (2).
Push the furnace back:
Lift the bayonet catch (2) slightly so as not to damage the seal on the combustion tube
connection.
Close the bayonet catch by turning the bayonet catch handle (1) to the left.
Close the ground-in clamp (3) on the Reductor again.
Connect the post combustion tube with the two ground-in clamps again.
Connect the O
2
supply tube with a ground-in clamp.
Close the front door.
Phase 4: Follow-up work, gas pressure/ tightness
After completing all maintenance work, put the instrument back into a state ready for measurement.
See "Follow-up preparation of the instrument: Gas pressure / tightness" in Preparing and following up
maintenance work.
To be observed after maintenance
If you have performed maintenance work that has a defined maintenance interval, you must manually
reset the "Status" field to zero on the "Maintenance intervals" dialog (accessible via Options >
Maintenance > Intervals. See Defining maintenance intervals in the software.
Содержание DuMaster D-480
Страница 1: ...DuMaster D 480 Operation Manual 11593626C en...
Страница 4: ......
Страница 12: ......
Страница 16: ......
Страница 30: ......
Страница 102: ......
Страница 181: ...Operating instructions DuMaster D 480 B CHI Labortechnik AG 8 Maintaining the instrument 181...
Страница 286: ...Operating instructions DuMaster D 480 B CHI Labortechnik AG 10 Menu and dialog descriptions 286...
Страница 338: ......
Страница 352: ......
Страница 353: ......