Operating instructions DuMaster D-480
©BÜCHI Labortechnik AG
9 - Repairing the instrument
249
If 21 CFR Part 11 functionality is enabled, the "Comment change" dialog is displayed. Enter a
justification in the "Comment modification" dialog and click
Close
.
Changing the position of the carousel
When is this necessary?
It is necessary to change the position of the carousel in the following cases:
If the analyzer was switched off, the carousel is no longer in zero position. Before you can start
an analysis, you must move the carousel back to zero position.
If you lose samples, it may be necessary to move the carousel to a different position.
Changing the position
Please observe the following instruction:
Caution
Inserting multiple samples into the ball valve (only for
solid operation)
If there are samples in the carousel while you are changing the position of the
carousel via
System > Carousel position
, multiple samples may fall into the ball
valve at the same time. In this case, a service technician is required.
Before changing the position of the carousel, always remove all samples from the
carousel.
Proceed as follows:
Remove all samples from the carousel.
Open the "Set carousel position" dialog via
System > Sample Position
.
Once you have removed all samples from the carousel, check the checkbox next to "All samples
removed from carousel".
Decide:
If you want to...
... move the carousel to a position other than zero position, select the radio button next to
"Position" and use the arrows to set the position in the box next to it.
... move the carousel to the zero position, select the radiobutton next to "Reference run".
Click
OK
. This closes the dialog and moves the carousel to the desired position.
Export analysis data for support
Export for support
The service department can draw conclusions about the condition of the analyzer from the analysis
data, particularly the graphical information. Therefore, analysis data must be exported in sqlite format
for support, as this also contains graphical data.
Exporting analysis data
Proceed as follows:
Open the "Export document" dialog in the operating software by selecting
File > Export/Import >
Export
.
Select the desired document file and click
OK
. This displays the "Save file as" dialog.
Select the directory to export the document to from the "Save" dropdown listbox.
Содержание DuMaster D-480
Страница 1: ...DuMaster D 480 Operation Manual 11593626C en...
Страница 4: ......
Страница 12: ......
Страница 16: ......
Страница 30: ......
Страница 102: ......
Страница 181: ...Operating instructions DuMaster D 480 B CHI Labortechnik AG 8 Maintaining the instrument 181...
Страница 286: ...Operating instructions DuMaster D 480 B CHI Labortechnik AG 10 Menu and dialog descriptions 286...
Страница 338: ......
Страница 352: ......
Страница 353: ......