Operating instructions DuMaster D-480
©BÜCHI Labortechnik AG
9 - Repairing the instrument
247
Insert additional sample lines at the top of the sample view for a daily factor determination via
Edit > Insert Line
.
Prepare daily factor samples and place them on the carousel. Put the other samples behind this.
Restart the measurement via
System > Continuous
.
Reacting to a power failure
Please observe the following instruction:
Caution
Inserting multiple samples into the ball valve (only for
solid operation)
If there are samples in the carousel while you are changing the position of the
carousel via
System > Carousel position
, multiple samples may fall into the ball
valve at the same time. In this case, a service technician is required.
Before changing the position of the carousel, always remove all samples from the
carousel.
Proceed as follows:
If the analyzer is still switched on, switch it off and wait for a moment.
Switch the analyzer on again.
If the ball valve...
... audibly does perform a reference run: Go to the next step.
... ...does not perform a reference run: Disconnect the instrument from the mains, check the
fuses and replace them if necessary. See Replacing fuses. After replacing the fuse repeat
this step.
Restart the operating software. The temporary file of the most recent document is loaded.
All measuring results exist except for the most recent sample. A message appears which informs
that the sample position is not definied and a reference run has to be carried out.
Confirm this message by clicking
OK
. This displays the "Adjusting the carousel position dialog.
without removing the carousel, remove all remaining samples from the carousel. Confirm this in
the dialog by checking "All samples removed from carousel". Then move the empty carousel to
zero position ("Reference run" radio button). Cf.
Changing the position of the carousel
(on
Create a backup of the current document via
File > Save as
.
Perform a leak test by selecting
Options > Diagnostics > Leak Test
.See Performing a leak test.
If a leak is identified, replace the sealing elements in the furnace area. See Replacing sealing
elements.
Correct the number of the current sample via
Edit > Current sample
. See Adjusting the current
sample.
Insert additional sample lines at the top of the sample view for a daily factor determination via
Edit > Insert Line
.
Prepare daily factor samples and place them on the carousel. Put the other samples behind this.
Restart the measurement via
System > Continuous
.
Checking air-tightness of instrument
Proceed as follows:
Perform a leak test to check the air-tightness of the instrument. See Performing a leak test.
If any leaks are detected, replace all sealing elements in the furnace area. See Replacing sealing
elements.
Содержание DuMaster D-480
Страница 1: ...DuMaster D 480 Operation Manual 11593626C en...
Страница 4: ......
Страница 12: ......
Страница 16: ......
Страница 30: ......
Страница 102: ......
Страница 181: ...Operating instructions DuMaster D 480 B CHI Labortechnik AG 8 Maintaining the instrument 181...
Страница 286: ...Operating instructions DuMaster D 480 B CHI Labortechnik AG 10 Menu and dialog descriptions 286...
Страница 338: ......
Страница 352: ......
Страница 353: ......