Operating instructions DuMaster D-480
©BÜCHI Labortechnik AG
9 - Repairing the instrument
240
Performing a leak test
Purpose of the leak test
The leak test checks the instrument for leaks.
Performing a leak test
The leak test consists of a total leakage test and a subsequent leak test stages.
The leak test stages checks the gas-tightness of the system by using severer criteria and, in case a
leak has been detected, supports you during the leak search procedure.
Proceed as follows:
Click
Options > Diagnosis > Leak test
. A message window appears which requests you to
confirm the execution of the leak test once more.
Click
Yes
to start the leak test. The program opens the "Leak test: Whole device" dialog. The
tubing diagram of the analysis system is displayed.
Click
Start
to begin "Whole System" leak test. The "Leak test: Whole device" dialog appears.
The leak test begins. On the left, the arrow indicates the current stage of the leak test.
The result of the leak test is displayed at the bottom:
If the leak test has passed the symbol
becomes visible.
If the leak test has failed the symbol
becomes visible.
Print out the result if you wish. Click
to do so.
Click
Close
to close the dialog again.
If the test for the whole system has not been passed, click now
Next
. This will activate the next
test stage. Here you will be informed how to proceed. Click
Start
to continue with the
corresponding test stage. The "Test Phases" dialog will open again showing the status of the
currently running test stage.
Note:
Switching to the next test stage is only possible during performing a fine leak test!
Continue with the stages described above until the result "Leak test not passed" is displayed.
The leak is then probably located at the blue indicated gas ways.
Further information
Leak test.
Leak test: Test phases
(on page 311).
Leak test procedure
(on page 241).
Содержание DuMaster D-480
Страница 1: ...DuMaster D 480 Operation Manual 11593626C en...
Страница 4: ......
Страница 12: ......
Страница 16: ......
Страница 30: ......
Страница 102: ......
Страница 181: ...Operating instructions DuMaster D 480 B CHI Labortechnik AG 8 Maintaining the instrument 181...
Страница 286: ...Operating instructions DuMaster D 480 B CHI Labortechnik AG 10 Menu and dialog descriptions 286...
Страница 338: ......
Страница 352: ......
Страница 353: ......