Operating instructions DuMaster D-480
©BÜCHI Labortechnik AG
8 - Maintaining the instrument
195
When to replace sealing elements
What sealing elements are used?
The following sealing elements are used:
Half shells
O-rings
Quad rings
Ferrules
When to replace sealing elements
The following table gives an overview of what instrument components contain which sealing elements
and when to replace these sealing elements:
Instrument component
What to replace
To be replaced when...
Ball valve
Half shells
... the ball valve is serviced and the half shells
exhibit grooves.
O-rings
... the ball valve is serviced and
the o-rings exhibit cracks or other damage
or
the o-rings no longer protrude round out of
the grooves.
Quartz glass bridge
O-rings
... you cannot feel any elastic resistance when
you press in the o-rings when installing the
quartz glass bridge.
Plug of Reductor
®
O-rings
... you can hardly feel any contact pressure
when inserting the Reductor plug.
... the o-rings no longer protrude round out
of the grooves of the plug.
Quad ring
... you cannot feel any elastic resistance when
pressing on the ball.
Drying tube (small, left)
O-rings
... the o-rings no longer protrude round out of
the grooves of the plug.
Combustion tube connection
with Ni flap
O-rings
... you can hardly feel any contact pressure
when inserting the combustion tube
connector into the combustion tube.
... the o-rings no longer protrude round out
of the grooves of the combustion tube
connector.
Post combustion tube
O-ring
... you can hardly feel any contact pressure
when inserting the post combustion tube
plug.
... the o-rings no longer protrude round out
of the grooves of the plug.
Quad ring
... you cannot feel any elastic resistance when
pressing on the ball.
Drying tube (large, right)
O-rings
... the o-rings no longer protrude round out of
the grooves of the plug.
Содержание DuMaster D-480
Страница 1: ...DuMaster D 480 Operation Manual 11593626C en...
Страница 4: ......
Страница 12: ......
Страница 16: ......
Страница 30: ......
Страница 102: ......
Страница 181: ...Operating instructions DuMaster D 480 B CHI Labortechnik AG 8 Maintaining the instrument 181...
Страница 286: ...Operating instructions DuMaster D 480 B CHI Labortechnik AG 10 Menu and dialog descriptions 286...
Страница 338: ......
Страница 352: ......
Страница 353: ......