8 - Maintaining the instrument
224
Fill in the quartz wool.
Fill in the Regainer.
Insert the plug at the top (filled with brass wool).
Reductor replacing
Note
The top and bottom plug of the Reductor are no consumables and therefore have to be reused.
Use both plugs for the new Reductor.
The Reductor is a consumable and should be replaced.