Operating instructions DuMaster D-480
©BÜCHI Labortechnik AG
10 - Menu and dialog descriptions
329
Sleep / wake-up function (Dialog)
Function
In this dialog you define the sleep and wake-up options for the analyzer.
Note:
Observe the following when entering the date in the "Wake-up" section: The program accepts
numbers between 1 and 31 for day input, but it does not check whether the date is permissible (e.g.
09/31).
Open
Open the dialog by selecting
Options > Settings > Sleep/Wake-up
.
Elements
The following describes the elements of the dialog.
In section "Sleep":
Label
Element type
Meaning
Sleep disabled
Radiobutton
The sleep/wake-up function is disabled.
Sleep after end of
sample
Radiobutton
The analyzer goes into sleep mode after finishing the last
sample (indicated by the stop marker).
Sleep at sample No. Option field /
textbox
The analyzer goes into sleep mode when it reaches the
specified sample.
You can enter the sample number in the adjacent textbox.
Shut off carrier gas
Checkbox
Shuts off the carrier gas when the analyzer goes into sleep
mode.
Switch of combustion
tube
Checkbox
The heater of the combustion tube will be turned off.
Switch of post
combustion tube
Checkbox
The heater of the combustion tube will be turned off.
Switch off the
Reductor
Checkbox
The heater of the Reductor will be turned off.
Sleep now
Button
Causes the analyzer to go into sleep mode immediately.
In section "Wake-up":
Label
Element type
Meaning
Time
Textbox
Fields to input the time in hours and minutes at which to
wake up the analyzer.
Date
Textbox
Fields to input the date on which to wake up the analyzer.
One-time wake-up at
above time/date
Radiobutton
Wakes up the sleeping analyzer at the date and time
specified above.
Daily wake-up at
above time/date,
except
Radiobutton
Wakes up the analyzer every day at the time specified
above.
Mon. Tues. Wed.
Thurs. Fri. Sat. Sun.
Checkbox
Defines on which days not to wake up the instrument, e.g.
on weekends.
Continue after wake-
up if there are
samples
Checkbox
If there are any unprocessed samples in sample memory,
analysis is automatically resumed after "wake-up".
Wake up now
Button
Wakes up the analyzer immediately.
Содержание DuMaster D-480
Страница 1: ...DuMaster D 480 Operation Manual 11593626C en...
Страница 4: ......
Страница 12: ......
Страница 16: ......
Страница 30: ......
Страница 102: ......
Страница 181: ...Operating instructions DuMaster D 480 B CHI Labortechnik AG 8 Maintaining the instrument 181...
Страница 286: ...Operating instructions DuMaster D 480 B CHI Labortechnik AG 10 Menu and dialog descriptions 286...
Страница 338: ......
Страница 352: ......
Страница 353: ......