Operating instructions DuMaster D-480
©BÜCHI Labortechnik AG
10 - Menu and dialog descriptions
326
Valid for all tabs:
Label
Element type
Meaning
Section
Dropdown listbox This field lists the available sections. Field to input a name for
a new section.
Entries
Listbox
This field lists the entries of a previously selected section.
New Section
Button
Creates a new section under the name previously entered in
the dropdown listbox.
Edit
Button
Opens the "Configure key value" dialog. Allows you to edit a
section entry previously selected from the list.
Delete
Button
Deletes a section previously selected in the dropdown listbox.
Deletes a section entry previously selected in the listbox.
Paste
Button
Opens the "Configure key value" dialog. Allows you to insert
new entries in a section.
Import
Button
Opens the "Open" dialog. Allows you to import further
sections.
Export
Button
Opens the "File save as" dialog. Allows you to export
sections.
Delete all
Button
Deletes all entries which are included in the activated tab.
OK
Button
Saves new settings
"DEVICE.INI" tab
The following table describes the entries under the "AnalyzerPort" section:
Entry
Standard
Description
Name
COM1
Defines the serial interface between the PC and the analyzer.
Baud
9600
Defines the transfer rate of the serial interface between the PC
and the analyzer.
Data
8
Defines the number of databits of the serial interface between the
PC and the analyzer.
Stop
1
Defines the number of stopbits of the serial interface between the
PC and the analyzer.
Parity
N
Defines the type of parity check of the serial interface between the
PC and the analyzer.
N= no parity check
E= even parity check
O= odd parity check.
The following table describes the entries under the "BalancePort" section:
Entry
Standard
Description
Name
----
Defines the serial interface between the PC and the balance.
Baud
300
Defines the transfer rate of the serial interface between the PC
and the balance.
Data
7
Defines the number of databits of the serial interface between the
PC and the balance.
Stop
1
Defines the number of stopbits of the serial interface between the
PC and the balance.
Содержание DuMaster D-480
Страница 1: ...DuMaster D 480 Operation Manual 11593626C en...
Страница 4: ......
Страница 12: ......
Страница 16: ......
Страница 30: ......
Страница 102: ......
Страница 181: ...Operating instructions DuMaster D 480 B CHI Labortechnik AG 8 Maintaining the instrument 181...
Страница 286: ...Operating instructions DuMaster D 480 B CHI Labortechnik AG 10 Menu and dialog descriptions 286...
Страница 338: ......
Страница 352: ......
Страница 353: ......