8 - Maintaining the instrument
175
Viewing calibration coefficients
Viewing calibration coefficients
Proceed as follows:
Open the "Calibration coefficients" dialog in the operating software by selecting
Calibration >
Coefficients
.
Highlight the name of a coefficient set in the listbox.
The calibration coefficients of the coefficient set and the limits the calibration is valid within are
displayed.
Optimizing instrument condition for calibration
Note
Note that calibration only makes sense if you have previously checked and optimized the condition of
the instrument.
Why does the condition of the instrument need to be checked and optimized?
The condition of the instrument needs to be checked and optimized in order to avoid...
... leaks of the instrument.
... old tube fillings from distorting results.
Checking and optimizing the condition of the instrument
Proceed as follows:
Replace all tube fillings. See
Emptying and filling reaction tubes.
Filling the drying tube
(on page 233).
Replace the ash crucible. See Replacing the ash crucible.
Perform a leak test. See Performing a leak test.
Determine the blank value as a measure of gas-tightness. Perform blank value determination
during calibration measurement.See Types of blank value determination and their settings.
Perform test measurements so that the instrument is stable when you subsequently perform the
calibration measurement. Measure 10 x 250 mg aspartic acid.
Performing the calibration
Target group
Personnel authorized by the customer. Successful calibration is only possible if the user is familiar
with the operating instructions and the analyzer.
Note
Perform calibration carefully as perfect calibration is a condition for correct measuring results.
Calibration must comprise the complete working range.
Содержание DuMaster D-480
Страница 1: ...DuMaster D 480 Operation Manual 11593626C en...
Страница 4: ......
Страница 12: ......
Страница 16: ......
Страница 30: ......
Страница 102: ......
Страница 181: ...Operating instructions DuMaster D 480 B CHI Labortechnik AG 8 Maintaining the instrument 181...
Страница 286: ...Operating instructions DuMaster D 480 B CHI Labortechnik AG 10 Menu and dialog descriptions 286...
Страница 338: ......
Страница 352: ......
Страница 353: ......