7 - Using the instrument
140
Display on the PC "Press. Output": 900 - 1000 mbar
Is the flow meter display OK?
Measuring gas flow meter on the PC "MFC CO2": 700 ml/min
CO
2
flow meter on the PC "Flow CO2": approx. 700 ml/min
O
2
flow meter on the PC during O
2
dosing "MFC O2": as selected in the method
Does the desiccant still have sufficient capacity?
At least the last third should still be unused.
Have the maintenance intervals been observed? Check:
The drying tube.
The Ash crucible / ash finger.
The Regainer.
The Reductor.
Is the detector signal small and stable?
Is the proper method allocated to the respective samples?
Is the carousel loaded?
If you answer yes to all of the above, you can start measuring blank value, conditioning and daily
factor samples.
Checklist for real sample measurement
Purpose
The checklist helps you to check whether the instrument is ready to measure real samples.
Checklist
Go through the following list:
Were the results of a blank value determination OK?
Were the results of the conditioning sample measurement OK?
Have you processed a series of measurements with daily factor samples?
Are the operating gases on?
Are the furnaces at setpoint temperature?
Is pressure OK?
Display on the PC "Press. Input": 1300 - 1350 mbar
Display on the PC "Press. Output": 900 - 1000 mbar
Is the flow meter display OK?
Measuring gas flow meter on the PC "MFC CO2": 700 ml/min
CO
2
flow meter on the PC "Flow CO2": approx. 700 ml/min
O
2
flow meter on the PC during O
2
dosing "MFC O2": as selected in the method
Does the desiccant still have sufficient capacity?
At least the last third should still be unused.
Have the maintenance intervals been observed? Check:
The drying tube.
Содержание DuMaster D-480
Страница 1: ...DuMaster D 480 Operation Manual 11593626C en...
Страница 4: ......
Страница 12: ......
Страница 16: ......
Страница 30: ......
Страница 102: ......
Страница 181: ...Operating instructions DuMaster D 480 B CHI Labortechnik AG 8 Maintaining the instrument 181...
Страница 286: ...Operating instructions DuMaster D 480 B CHI Labortechnik AG 10 Menu and dialog descriptions 286...
Страница 338: ......
Страница 352: ......
Страница 353: ......