Operating instructions DuMaster D-480
©BÜCHI Labortechnik AG
7 - Using the instrument
132
Open the "Open" dialog by selecting
File > Export/Import > Import
.
Select the "Balance Files (*.dat)" file type from the "File type" dropdown listbox.
From the dropdown listbox select the directory containing the weighing data file.
Select a file and click
Open
.
Decide:
If you import into...
... an empty sample memory: The weighing data are saved to sample memory starting at
No. 1. Save the document. The import is finished.
... a sample memory that is not empty: The "Reading Balance file" dialog is displayed. Go to
the next step.
Select the desired radiobutton in the dialog:
If you...
... select "Append samples as from the current weighed sample": The weighing data are
stored starting after the current sample weight.
... select "Create new document": The weighing data are saved in a new document.
Click
OK
. The weighing data are imported and displayed in the sample field.
Save the document. The import is finished.
Prioritizing urgent samples
Prioritizing urgent samples
Proceed as follows:
Open the "Swap samples" dialog in the operating software by selecting
Edit > Swap
.
Enter the numbers of the samples to swap in the textboxes in the dialog. Then click
OK
.
This closes the dialog and the samples are swapped in the sample memory.
Manually swap samples in the carousel.
Optimizing sleep and wake-up behavior
What can you optimize?
You can set sleep/wake-up behavior so that conditioning samples in a series are measured before
the first user starts work on the instrument in the morning.
Note
Remember to prepare the appropriate conditioning samples and to sort them into the carousel after
the samples of the last measurement of the day.
Optimizing sleep and wake-up behavior
Proceed as follows:
Open the "Sleep/wake-up functions" dialog in the operating software by selecting
Options >
Settings > Sleep/Wake-up
.
The last measurement of the day must be stopped before the appended conditioning samples.
To do so, select the radio button next to "Sleep at sample No." in the "Sleep" section. In the box
next to it, enter the sample number at which you want the instrument to sleep; it corresponds to
the first conditioning sample.
Содержание DuMaster D-480
Страница 1: ...DuMaster D 480 Operation Manual 11593626C en...
Страница 4: ......
Страница 12: ......
Страница 16: ......
Страница 30: ......
Страница 102: ......
Страница 181: ...Operating instructions DuMaster D 480 B CHI Labortechnik AG 8 Maintaining the instrument 181...
Страница 286: ...Operating instructions DuMaster D 480 B CHI Labortechnik AG 10 Menu and dialog descriptions 286...
Страница 338: ......
Страница 352: ......
Страница 353: ......