Operating instructions DuMaster D-480
©BÜCHI Labortechnik AG
10 - Menu and dialog descriptions
303
Label
Element type
Meaning
Cancel
Button
Discards calibration coefficients that have just been
computed or input.
Further information
Further information includes instructions in which the dialog is used and knowledge articles
concerning the dialog:
Viewing calibration coefficients
(on page 175).
Defining standard substances as calibration samples
(on page 120).
Performing calibration.
Background knowledge required for calibration
(on page 51).
Configure calibration (Dialog)
Function
In this dialog you can specify the calibration curve calculation methods. It is not possible to perform
the calibration computation without this information.
Open
Open the dialog via
Calibration > Calibrate
. In order to open the dialog, sample memory must
contain a sufficient number of valid calibration samples.
Elements
The following describes the elements of the dialog.
On the respective tab:
Label
Element type
Meaning
One calibration range Radiobutton
Only one calibration curve is computed over the entire
measuring range for the respective element.
Polynomial degree
Textbox
Field to input the polynomial degree for the calibration
curve calculation over the entire measuring range. May be
set from 1 to 4.
Two calibration
ranges
Radiobutton
For each element, the whole measuring range is divided
into an upper and lower calibration range. A separate
calibration curve is computed for each calibration range.
Polynomial degree,
lower range
Textbox
Field to input the polynomial degree for the calibration
curve calculation in the lower range. May be set from 1 to
4.
Polynomial degree,
upper range
Textbox
Field to input the polynomial degree for the calibration
curve calculation in the upper range. May be set from 1 to
4.
Remaining elements:
Label
Element type
Meaning
OK
Button
Continues with the calibration calculation. If the option
"Two calibration ranges" has been checked for at least for
one element, then the "Specify lower and upper calibration
range" dialog will open. Here, the split-up of both
calibration ranges can be carried out.
Cancel
Button
Aborts the process.
Содержание DuMaster D-480
Страница 1: ...DuMaster D 480 Operation Manual 11593626C en...
Страница 4: ......
Страница 12: ......
Страница 16: ......
Страница 30: ......
Страница 102: ......
Страница 181: ...Operating instructions DuMaster D 480 B CHI Labortechnik AG 8 Maintaining the instrument 181...
Страница 286: ...Operating instructions DuMaster D 480 B CHI Labortechnik AG 10 Menu and dialog descriptions 286...
Страница 338: ......
Страница 352: ......
Страница 353: ......