Operating instructions DuMaster D-480
©BÜCHI Labortechnik AG
10 - Menu and dialog descriptions
307
Maintenance intervals (Dialog)
Function
In this dialog, any number of maintenance events can be defined and viewed.
Open
Open the dialog via
Options > Maintenance > Intervals
.
Elements
The following describes the elements of the dialog.
Label
Element type
Meaning
---
Listbox
In this field the texts of all currently defined events are
displayed. Each text has a highlighted bar which displays
the urgency of the maintenance:
Green bar: There is still time for maintenance.
Yellow bar: The maintenance must be performed
soon.
Red bar: The maintenance is overdue.
Total:
Label
Displays the number of all analyses performed so far.
Event: {n}
Textbox/label
Field to enter the text of a new event.
or
The field displays the text of the event selected in the
adjacent list box.
Interval
Textbox/label
Field to input the number of analyzed samples after
which the maintenance must be performed.
or
The field displays the interval of the event selected in
the adjacent list box.
Status
Textbox/label
Field to input the number of samples already analyzed,
e.g. when you install a used tube. You must enter 0 as
the status after performing maintenance.
or
This field displays the number of samples already
analyzed for the adjacent event.
Mode assignment
List box with
checkboxes
Allows assignment of the corresponding event to the
different analysis modes. Incrementing of the
corresponding counter will be suppressed if analysis
modes are not checked here.
Information only
Radiobutton
When a maintenance event falls due, the system only
displays a message in the status view. See Defining
maintenance intervals in the software.
Information and stop
continuous analysis
Radiobutton
When a maintenance event falls due, the system displays a
message in the status view and continuous analysis is
stopped.
New
Button
Allows the definition of a new event.
Delete
Button
Removes the selected event from the list.
Save
Button
Saves the settings made.
Close
Button
Closes the dialog.
Содержание DuMaster D-480
Страница 1: ...DuMaster D 480 Operation Manual 11593626C en...
Страница 4: ......
Страница 12: ......
Страница 16: ......
Страница 30: ......
Страница 102: ......
Страница 181: ...Operating instructions DuMaster D 480 B CHI Labortechnik AG 8 Maintaining the instrument 181...
Страница 286: ...Operating instructions DuMaster D 480 B CHI Labortechnik AG 10 Menu and dialog descriptions 286...
Страница 338: ......
Страница 352: ......
Страница 353: ......