Operating instructions DuMaster D-480
©BÜCHI Labortechnik AG
10 - Menu and dialog descriptions
291
In the dialog:
you are warned about adjusting the carousel position if the carousel contains samples,
you can move the carousel to zero position or any other position.
Please observe the following instruction:
Caution
Inserting multiple samples into the ball valve (only for
solid operation)
If there are samples in the carousel while you are changing the position of the
carousel via
System > Carousel position
, multiple samples may fall into the ball
valve at the same time. In this case, a service technician is required.
Before changing the position of the carousel, always remove all samples from the
carousel.
Note
Observe the following when adjusting the position of the carousel:
The carousel is moved so that the selected position is above the insertion opening. When the next
analysis starts, the carousel moves the next sample position above the insertion opening and drops
the sample. Therefore, it makes sense to select the previous position when moving the carousel.
Example: You want to move the carousel so as to analyze the sample at position 26 in the next
analysis. Therefore, select position 25. The carousel moves so far that position 25 is above the
insertion opening. When the next analysis starts, the carousel moves one position on during the
insertion process and thus drops the samples at position 26.
Open
Open the dialog via
System > Sample Position
.
Elements
The following describes the elements of the dialog.
Label
Element type
Meaning
All samples removed
from the carousel
Checkbox
If you have removed all samples from the carousel, you
must check this box. Otherwise the carousel will not move.
Position
Radiobutton
Moves the carousel to the desired position. You can set the
position number in the textbox to the right. You must input
the desired position number minus 1.
Reference run
Radiobutton
The carousel performs reference run, i.e. the carousel first
moves to a defined reference position and then to position
0 which has almost the same location.
OK
Button
Adjusts the position of the carousel, if confirmed above that
all samples have been removed from the carousel.
Cancel
Button
Aborts the process.
Further information
Further information includes instructions in which the dialog is used and knowledge articles
concerning the dialog:
Performing measurement work
(on page 134).
Reacting to a power failure
(on page 245).
Performing calibration.
Содержание DuMaster D-480
Страница 1: ...DuMaster D 480 Operation Manual 11593626C en...
Страница 4: ......
Страница 12: ......
Страница 16: ......
Страница 30: ......
Страница 102: ......
Страница 181: ...Operating instructions DuMaster D 480 B CHI Labortechnik AG 8 Maintaining the instrument 181...
Страница 286: ...Operating instructions DuMaster D 480 B CHI Labortechnik AG 10 Menu and dialog descriptions 286...
Страница 338: ......
Страница 352: ......
Страница 353: ......