Operating instructions DuMaster D-480
©BÜCHI Labortechnik AG
8 - Maintaining the instrument
227
Top view furnace
The following picture shows a top view of the furnace pulled out:
Legend:
1 Post combustion tube plug
2 Post combustion tube
3 Combustion tube connection
4 Combustion tube
5 Plug of Reductor
6 Reductor
Phase 2: Removing reaction tubes, continued
Proceed as follows:
Press the reductor plug (5), the combustion tube connection (3) and the post combustion tube
connector (1) down lightly and pull the furnace out of the instrument to the front.
Caution:
Tilting and jamming the combustion tube connection can damage the combustion tube.
Pull the combustion tube connection (3) out of the combustion tube with careful jiggling motions.
Place the combustion tube connection (3) on the tube rack.
Содержание DuMaster D-480
Страница 1: ...DuMaster D 480 Operation Manual 11593626C en...
Страница 4: ......
Страница 12: ......
Страница 16: ......
Страница 30: ......
Страница 102: ......
Страница 181: ...Operating instructions DuMaster D 480 B CHI Labortechnik AG 8 Maintaining the instrument 181...
Страница 286: ...Operating instructions DuMaster D 480 B CHI Labortechnik AG 10 Menu and dialog descriptions 286...
Страница 338: ......
Страница 352: ......
Страница 353: ......