7 - Using the instrument
121
Specifying the computation method for blank value and daily factor
Defining the blank value calculation method
Proceed as follows:
Open the "Configure Calculations" dialog in the operating software by selecting
Options >
Settings > Calculation
.
In section "Blank value determination" define the calculation method:
If you...
... select "sequentially" a separate blank value is computed for every element in every group
of blank value samples; this value is entered in the respective blank value column until the
next group of blank value samples.
... select "entirely" all blank value samples in the series are used to compute the blank
value. This blank value is then applied to all samples in the series.
... select "manual input" blank value calculation is not performed - even if there are blank
samples in the series. This allows manual input of blank values for individual samples.
... "autom. activate "automatic blank value calculation" the blank value is automatically
computed at the end of every analysis. This option ist not available if manual input has
been checked.
Then click
OK
to save your settings.
Defining the daily factor calculation method
Proceed as follows:
Open the "Configure Calculations" dialog in the operating software by selecting
Options >
Settings > Calculation
.
In section "Factor determination" define the calculation method:
If you...
... select "sequentially" then the daily factor is computed group-by-group and changes
throughout the day.
... select "entirely" the daily factor is computed on the basis of all factor samples in the file.
An average factor is computed.
... select "manual input" factor calculation is not performed - even if there are factor samples
in the series. This allows manual input of daily factors for individual samples.
... "autom. activate "automatic factor calculation" the factor is automatically computed at the
end of every analysis. This option ist not available if manual input has been checked.
... activate "upper/lower calib. range distinction" both for the upper and lower calibration
range separate factors will be calculated. This requires separate factor samples for each
range.
Then click
OK
to save your settings.
Enabling/disabling acoustic signals
Enabling/disabling acoustic signals
Proceed as follows:
Open the "Acoustic signals" dialog by selecting
Options > Settings > Acoustic Signals
.
Decide:
If you want to...
... switch on a signal, check the appropriate checkbox.
Содержание DuMaster D-480
Страница 1: ...DuMaster D 480 Operation Manual 11593626C en...
Страница 4: ......
Страница 12: ......
Страница 16: ......
Страница 30: ......
Страница 102: ......
Страница 181: ...Operating instructions DuMaster D 480 B CHI Labortechnik AG 8 Maintaining the instrument 181...
Страница 286: ...Operating instructions DuMaster D 480 B CHI Labortechnik AG 10 Menu and dialog descriptions 286...
Страница 338: ......
Страница 352: ......
Страница 353: ......