Operating instructions DuMaster D-480
©BÜCHI Labortechnik AG
8 - Maintaining the instrument
176
Process
The section is divided into the following phases:
Prepare samples and perform measurement
Ascertaining appropriate calculation methods
Set calculation methods, perform calculation
Optimizing calibration curves and apply calibration
Requirements:
Before starting work, the following requirements must be met:
Read
Background knowledge required for calibration
(on page 51) and
Assessing the
calibration curves
(on page 179).
Prepare a suitable balance. A reading precision of 0.01 mg is usually sufficient.
If you have not already done so, define the following by selecting
Options > Settings >
Standards
...
... the required calibration substances. See
Defining standard substances as calibration
samples
(on page 120).
... the required substances for daily factor determination. See
Defining standard
substances as factor and monitor samples
(on page 120).
Check and optimize the condition of the instrument. Otherwise it makes no sense to calibrate.
See Optimizing instrument condition for calibration.
Phase 1: Prepare samples and perform measurement
Please observe the following instruction:
Caution
Inserting multiple samples into the ball valve (only for
solid operation)
If there are samples in the carousel while you are changing the position of the
carousel via
System > Carousel position
, multiple samples may fall into the ball
valve at the same time. In this case, a service technician is required.
Before changing the position of the carousel, always remove all samples from the
carousel.
Proceed as follows:
Move the empty carousel to zero position. To do so, open the "Set carousel position" dialog in
the operating software by selecting
System > Sample Position
.
Select the radio button next to "Reference run" and click
OK
. The carousel moves to zero
position.
Prepare the samples. Cf.
Preparing samples
(on page 123).
Prepare the samples for determining the daily factor. Use 3 x aspartic acid at 250 mg with
method "Standard".
Prepare conditioning samples and calibration samples. Refer to Calibration tables for the
recommended settings for each mode.
Insert the samples on the carousel.
Enter the following sample data in the sample view (text view) on the PC:
weight in [mg] if it has not been transmitted by the balance,
the name of the sample (max. 32 characters):
"Blnk" for blank samples,
Daily factor or calibration samples must have the name defined in
Options > Settings >
Standards
.
Select an appropriate method for run-in and calibration samples according to the following list:
Содержание DuMaster D-480
Страница 1: ...DuMaster D 480 Operation Manual 11593626C en...
Страница 4: ......
Страница 12: ......
Страница 16: ......
Страница 30: ......
Страница 102: ......
Страница 181: ...Operating instructions DuMaster D 480 B CHI Labortechnik AG 8 Maintaining the instrument 181...
Страница 286: ...Operating instructions DuMaster D 480 B CHI Labortechnik AG 10 Menu and dialog descriptions 286...
Страница 338: ......
Страница 352: ......
Страница 353: ......