8 - Maintaining the instrument
199
Furnace area
The following picture shows the details of the furnace area:
Legend:
1 Bayonet catch
2 Handle
3 Connection Reductor
4 Ball valve fastening
5 Ground-in clamp
6 O
2
supply tube
7 Post combustion tube connection
Phase 2: Pulling the furnace out of the instrument
Proceed as follows:
Open the front door.
Unscrew the combustion tube connector. To do so, turn the bayonet catch handle (2) to the right.
Remove the ground-in clamp (5) on the reductor (3).
Remove the ground-in clamp (5) on the post combustion tube (7).
Remove the O
2
supply tube (6) by opening the two ground-in clamps (5).
Top view furnace
The following picture shows a top view of the furnace pulled out:
Legend:
1 Post combustion tube plug
2 Post combustion tube
3 Combustion tube connection
4 Combustion tube
5 Plug of Reductor
6 Reductor
Phase 3: Removing the ball valve
Proceed as follows:
Open the ground-in ball-and-socket joints of the tube (1) leading to the ball valve.
Push the right side door open from inside (magnetic lock).
From the right, unscrew the ball valve (3) locking screws using an 8mm open jawed wrench.
Pull the ball valve out of the instrument to the front.
Содержание DuMaster D-480
Страница 1: ...DuMaster D 480 Operation Manual 11593626C en...
Страница 4: ......
Страница 12: ......
Страница 16: ......
Страница 30: ......
Страница 102: ......
Страница 181: ...Operating instructions DuMaster D 480 B CHI Labortechnik AG 8 Maintaining the instrument 181...
Страница 286: ...Operating instructions DuMaster D 480 B CHI Labortechnik AG 10 Menu and dialog descriptions 286...
Страница 338: ......
Страница 352: ......
Страница 353: ......