10 - Menu and dialog descriptions
290
System > Wake up (Command)
What is the command used for?
This command wakes up the "sleeping" system. The carrier gas is reactivated and the furnaces are
heated back up to their working temperatures. This command is used to manually wake up the
system if you need to start up the system prematurely, i.e. earlier than the chosen wake-up time.
Short cuts
Invocation via the toolbar:
Note
If you have defined a wake-up time that is in the past, you must wake up the sleeping system again
with this command.
Feeding (Dialog)
Function
In this dialog you specify the size of the sample carousel. Make sure that the setting here matches to
the actual size of the sample carousel.
Open
Open the dialog by selecting
System > Feeding
.
Elements
The following describes the elements of the dialog.
Label
Element type
Meaning
Carousel with {n}
positions
Radiobutton
The feeding system is switched to a carousel size of {n}
positions.
OK
Button
Saves the setting.
Cancel
Button
Aborts the process.
Further information
Further information includes instructions in which the dialog is used and knowledge articles
concerning the dialog:
Stopping continuous analysis
(on page 248).
Right mouse button function
(on page 67).
Adjust carousel position (Dialog)
Note
This function is only available in the solids modes.
Function
After a power failure or if you repeat initial startup and remove the carousel, it is necessary to readjust
the position of the carousel.
Содержание DuMaster D-480
Страница 1: ...DuMaster D 480 Operation Manual 11593626C en...
Страница 4: ......
Страница 12: ......
Страница 16: ......
Страница 30: ......
Страница 102: ......
Страница 181: ...Operating instructions DuMaster D 480 B CHI Labortechnik AG 8 Maintaining the instrument 181...
Страница 286: ...Operating instructions DuMaster D 480 B CHI Labortechnik AG 10 Menu and dialog descriptions 286...
Страница 338: ......
Страница 352: ......
Страница 353: ......