Operating instructions DuMaster D-480
©BÜCHI Labortechnik AG
10 - Menu and dialog descriptions
289
Note
The "Abort analysis" dialog appears.
Observe the notes listed therein. Abort analysis only if either the sample has not been thrown in yet or
has been already completely combusted. In the latter case, make sure that the column does not
retain any element components, i.e. the desorption procedure has been completed. Otherwise, an
abortion leads to system contaminations combined with significant damages.
To abort a running analysis, you have to confirm the note displayed in the dialog first. Only after the
confirmation the
Abort analysis
button becomes accessible, thus allowing the abortion.
System > Auto zero (Command)
What is the command used for?
This command causes to set the current output values of the dectors to zero. The purpose is to reset
the signals to zero if they have drifted over time.
Furnace switch off/on (Dialog)
Function
In this dialog the furnace will be switched off and on.
Open
Open the dialog via
System > Furnace
.
Elements
The following describes the elements of the dialog.
Label
Element type
Meaning
Yes
Button
Confirm the action.
No
Button
Aborts the process.
Initialize the instrument (Dialog)
Function
In this dialog the instrument will be initialized. It will be set all parameters to default values.
Open
Open the dialog via
System > Initialise
.
Elements
The following describes the elements of the dialog.
Label
Element type
Meaning
Yes
Button
Confirm the action.
No
Button
Aborts the process.
Содержание DuMaster D-480
Страница 1: ...DuMaster D 480 Operation Manual 11593626C en...
Страница 4: ......
Страница 12: ......
Страница 16: ......
Страница 30: ......
Страница 102: ......
Страница 181: ...Operating instructions DuMaster D 480 B CHI Labortechnik AG 8 Maintaining the instrument 181...
Страница 286: ...Operating instructions DuMaster D 480 B CHI Labortechnik AG 10 Menu and dialog descriptions 286...
Страница 338: ......
Страница 352: ......
Страница 353: ......