6 - Starting up and shutting down the instrument
111
Purpose of the transport protections
The transport protections prevent...
... the furnace section from slipping on its rails while the analyzer is being transported.
... the loosely attached water trap from damages while the analyzer is being transported.
Removing the furnace transport protection
Proceed as follows:
Open the front door of the instrument.
Push the right side door open from inside (magnetic lock). The transport protection is now visible
in the form of a bracket (1).
Unscrew the screws and remove the bracket.
Fasten the bracket on the rear casing angle (1). It should be stored there so it does not get lost.
Close the right side door and screw the big screw back in that keeps the door closed.
Removing the water trap transport protection
Proceed as follows:
Open the front door of the instrument.