19-11
IM DL950-02EN
Date: The file name is the date and time (down to ms) when the file is saved. The file
name specified with the File Name setting is not used.
Year
Month
Day
Hour
Minute
Second
ms
20200930_121530_100_000 (2020/09/30 12:15:30.100)
Serial number (000 to 999) that is appended when
a single file exceeds 2 GB
Regardless of whether the auto naming feature is set to OFF, Numbering, or Date, if the data
size of a single file exceeds 2 GB, an underscore and a three-digit sequence number (000 to
999) is appended to the file name. The sequence number is incremented by one each time a
file is added. This is appended only if the file exceeds 2 GB.
Setting a Comment (Comment)
You can add a comment that consists of up to 120 characters when you save a file. You do not
have to enter a comment. All characters, including spaces, can be used in comments.
Selecting the Data Type (Data Type)
3.
Tap
Data Type
, or press the soft key. The following screen appears.
When Dual Capture Mode Is Off
When Dual Capture Mode Is On
Saves in binary
format
Saves in ASCII
format
Saves in MATLAB
format
Saves low-speed sampling and
high-speed sampling waveform data to
a file in binary format
Saves high-speed sampling waveform
data in binary format
Saves low-speed sampling waveform data
in ASCII format
Saves low-speed sampling waveform
data in MATLAB format
Saves high-speed sampling waveform
data in ASCII format
Saves high-speed sampling waveform data
in MATLAB format
You can save the sampling data stored in the acquisition memory in the following data formats.
Data format Extension
Binary
.WDF
Data is saved in binary format.
You can load this type of data into the instrument. ► sec. 19.7
Data can be saved in compressed form. For details, see the next page.
ASCII
.CSV
Data is scaled using the specified range and saved to a file in ASCII
format.
You cannot load this type of data into the instrument.
MATLAB
.MAT
Data is saved in MATLAB format.
You cannot load this type of data into the instrument.
Waveforms measured using the dual capture feature cannot be saved.
Note
If the data format is ASCII or MATLAB and the combination of the record length and the number of channels
causes the file size to exceed 2 GB, the file cannot be saved.