11-5
IM DL950-02EN
Setting the Measurement Time Period (Time Range1/Time Range2)
The specified parameters are measured in the range between the measurement start point (Time
Range1) and measurement end point (Time Range2).
Note
For SSD recording waveforms, up to 100 Mpoint from the measurement start point (Time Range1) are
measured.
Setting the Time Measurement Reference Level and 1-Cycle Mode
(Detail)
2.
Tap
Detail
, or press the soft key. The following menu appears.
Set the distal value.
Set the proximal value.
Set the unit for the reference level
(% or Unit).
Set the mesial value.
Set the mode for determining high and
low levels (Auto, Max-Min).
Turns 1-cycle mode on or off
Set the target waveform.*
(Slide to display waveforms that do not fit in the screen.)
* Depending on the channel you select, you need to select sub channels and bits.
Note
When 1-cycle mode is on, if the interval between Time Range1 and Time Range2 is less than 1 cycle, the
measured value will be “*****.”
Setting the Indicator (Indicator)
The measurement area of the specified item is displayed with cursors.
2.
Tap
Indicator
, or press the soft key. The following menu appears.
Set the parameters you want to display.
(Select from the measurement source items.)
Set the reference waveform.*
(Slide to display waveforms that do not fit in the screen.)
* Depending on the channel you select, you need to select sub channels and bits.
The measurement areas of the following items can be indicated.
Peak to Peak, Amplitude, Maximum, Minimum, High, Low, Average, Middle, RMS, Std,
Deviation, +Overshoot, ‒Overshoot, Rise, Fall, Frequency, Period, +Width, ‒Width, Duty,
Burst1, Burst2, Avg Frequency, Avg Period, I, Integ2TY, Delay
11.1 Automatically Measuring Waveform Parameters