4-22
IM DL950-02EN
Setting the State and Trigger Conditions (Pattern Setup)
2.
Tap
Pattern Setup
, or press the soft key. The following menu appears.
On a module with logic inputs
or sub channels, expand the
menu, and set each bit or sub
channel.
Set the state condition (H, L, or X (do not use as a trigger source)).
Set the trigger level.
Set the hysteresis ( , , ).
Set the determination mode.
Set the reference time.*
* Set T1 and T2 when the determination mode is B Between.
Set Time when the determination mode is B < Time, B >Time, or B TimeOut.
Note
If linear scaling is applied to the trigger source, set the trigger level using a linearly scaled value.
Setting the Determination Mode (Mode)
Set what kind of relationship must be established between the state condition B achievement
time and the specified reference times (Time or T1 and T2) for the instrument to trigger.
B < Time
The instrument triggers when the achievement time is shorter than the reference time
(Time), and the state condition changes from being met to not being met.
B > Time
The instrument triggers when the achievement time is longer than the reference time
(Time), and the state condition changes from being met to not being met.
B TimeOut The instrument triggers when the achievement time is longer than the reference time
(Time).
B Between The instrument triggers when the achievement time is longer than reference time T1
and shorter than reference time T2, and the state condition changes from being met to
not being met.
4.13 Triggering on a Pulse Width Trigger (Enhanced)