![YOKOGAWA ScopeCorder DL950 User Manual Download Page 100](http://html.mh-extra.com/html/yokogawa/scopecorder-dl950/scopecorder-dl950_user-manual_911350100.webp)
4-12
IM DL950-02EN
Setting the State and Trigger Conditions (Pattern Setup)
2.
Tap
Pattern Setup
, or press the soft key. The following menu appears.
On a module with logic inputs or sub
channels, expand the menu, and set
each bit or sub channel.
Set the state condition (H, L, or X (do not use as a trigger source)).
Set the trigger level.
Set the hysteresis ( , , ).
Set the state condition
achievement conditions
(Enter, Exit).
Set the number of times state
condition B must be met.
Note
If linear scaling is applied to the trigger source, set the trigger level using a linearly scaled value.
4.8 Triggering on an A to B(N) Trigger (Enhanced)