13-9
IM DL950-02EN
Notes on Setting User-Defined Math Functions
When setting a user-defined math function, keep the following in mind.
• A math channel with a larger number cannot be used in an expression of a mass channel with a
smaller number.
• An expression with only constants cannot be defined.
• An expression can only contain up to two FILT1 or FILT2 functions.
• Other computations cannot be performed on a pulse width computation.
• When performing shift computation, pulse width computation, or binary conversion, only a single
operator can be used in an expression.
For details, see the Features Guide (IM DL950-01EN) contained in the CD.
13.4 Performing User-Defined Math (/G02 option)