![YOKOGAWA ScopeCorder DL950 User Manual Download Page 97](http://html.mh-extra.com/html/yokogawa/scopecorder-dl950/scopecorder-dl950_user-manual_911350097.webp)
4-9
IM DL950-02EN
4.6 Triggering on a Power Line Signal (Simple)
This section explains the settings that are used when triggering on a power line signal.
Applicable in Scope Mode
• Trigger source
► “Simple Trigger (Simple)” and “Trigger Source (Source)” in the Features Guide
Trigger Menu
1.
On the waveform screen’s
MENU
, tap
TRIGGER
>
MENU
, or press
TRIG.MENU
. A trigger
menu appears.
Select Simple.
Set the trigger source to Line.
Set the trigger position, trigger delay, and hold-off
► sec. 4.2
Set the trigger mode. ► sec. 4.1