3-10
IM DL950-02EN
Capture Mode (Capture Mode)
Set when to start or stop waveform acquisition when the START/STOP button is pressed.
There are two capture modes: Auto and On Start.
Auto (Auto): Low-speed sampling and high-speed sampling waveforms are aquired from
when acquisition is started until it is stopped. In high-speed sampling, when
the acquirable history waveform data is acquired, old history waveforms are
overwritten in order.
On Start (On Start): In low-speed sampling, when the specified record length of data or 10
divisions of data is acquired after starting waveform acquisition, waveform
acquisition stops. In high-speed sampling, when the acquirable history
waveform data is acquired, waveform acquisition stops.
Display Format (Format)
1.
With dual capture turned on, on the waveform screen’s
MENU
, tap
DISPLAY
>
DISPLAY
, or
press
DISPLAY
. The Display menu appears.
2.
Tap
Format
, or press the soft key.
Display format (number divisions) of
low-speed sampling waveforms
Set the display ratio of low-speed
sampling waveforms.
Display format (number divisions) of
high-speed sampling waveforms
3.
Set the display format (number divisions) for low-speed sampling and high-speed sampling.
For details on the display format and other settings, see chapter 5, “Display.”
Configuring the Waveform Display Area
2.
Tap
Low Speed
, or press the soft key.
3.
Set the percentage of the entire waveform display area that the low-speed sampling waveform
will occupy.
3.4 Using the Dual Capture Feature (Scope mode only)