![YOKOGAWA ScopeCorder DL950 User Manual Download Page 136](http://html.mh-extra.com/html/yokogawa/scopecorder-dl950/scopecorder-dl950_user-manual_911350136.webp)
6-8
IM DL950-02EN
Measuring X-Y Waveforms with Horizontal and Vertical Cursors (H
& V)
3.
Tap
Type
, or press the soft key to select
H & V
.
See “Setting Measurement Items.”
Set the vertical cursor position.*
You can also drag the cursors on the X-Y screen.
Select the trace to measure with cursors
(XY1 to XY8).
Set the cursor type to H & V.
Set Mode to ON.
Set the horizontal cursor position.*
You can also drag the cursors on the X-Y screen.
* Tap the cursor and then tap the arrow that appears (shown in the figure below)
to adjust the cursor position in fine increments.
Setting Measurement Items (Item Setup)
4.
Tap
Item Setup
, or press the soft key. The following screen appears.
Select the measurement items that you want to use.
X1:
Horizontal (X axis) value at V-Cursor1
X2:
Horizontal (X axis) value at V-Cursor2
Y1:
Vertical (Y axis) value at H-Cursor1
Y2:
Vertical (Y axis) value at H-Cursor2
ΔX:
Difference between the horizontal (X axis)
values of V-Cursor1 and V-Cursor2
ΔY:
Difference between the vertical (Y axis) values
of H-Cursor1 and H-Cursor2
ΔX/ΔY: Amount of change on the horizontal axis
corresponding to the amount of change on the
vertical axis
ΔY/ΔX: Amount of change on the vertical axis
corresponding to the amount of change on the
horizontal axis
6.3 Measuring X-Y Waveforms with Cursors