2-48
IM DL950-02EN
Setting the SENT Format (SENT Format Setup)
4.
Tap
SENT Format Setup
, or press the soft key. The following screen appears.
Turns Fast Channel Multiplexing on or off
You can set this when the SENT monitor module version is 0x07 or later.
Set the Slow CH message format.
(Short(ID 4bit + Data 8bit), Enhanced(ID 8bit +
Data 12bit), Enhanced(ID 4bit + Data 16bit))
Set the clock tick (1.00 to 100.00 µs).
Set the number of data nibbles of Fast
CH messages (1 to 8).
Set whether to include pause pulses
in Fast CH messages (ON or OFF).
Set the CRC method (Legacy, Recommended).
Setting Error Channels (Error Channel Setup)
4.
Tap
Error Channel Setup
, or press the soft key. The following screen appears.
Turns error detection on or off (Successive Calibration Pulses
(Option2) only)
Error detection (Detect) is always on for Fast Channel CRC, Slow
Channel CRC, Nibble Value, and Pulse Number.
Turns error trigger display on or off
Turns error count integration on or off
Turns error count reset on start on
or off
Executes the manual resetting of
the error count
Input Setup (Input Setup)
4.
Tap
Input Setup
, or press the soft key. The following screen appears.
Set the probe (1:1, 10:1).
Set the timeout value (0.1 to 2000.0ms).
Upper threshold (fixed)
Lower threshold (fixed)
2.11 Configuring the Monitoring of SENT Signals (/VCE option)