![YOKOGAWA ScopeCorder DL950 User Manual Download Page 147](http://html.mh-extra.com/html/yokogawa/scopecorder-dl950/scopecorder-dl950_user-manual_911350147.webp)
8-3
IM DL950-02EN
Setting Search Conditions (Setup)
3.
Press the
Setup
soft key. A menu appears according to the waveform to be searched you
specified.
When the waveform to search is set to CH1 to CH16, 16CH TEMP/VOLT,
CAN,* CAN FD,* LIN,* SENT*
Set the waveform to search (CH1 to CH16, 16CH TEMP/VOLT, CAN, CAN
FD, LIN, SENT).
Set the level.
Set the edge polarity ( , , ).
Set the hysteresis ( , , ).
Set the number of times to detect
the conditions.
* Use this menu for waveforms whose sub channel Value Type on the CAN bus monitor, CAN/CAN FD
monitor, or CAN & LIN bus monitor is Unsigned, Signed, or Float and waveforms whose sub channel on the
SENT monitor is FastCH, SlowCH, or Error Count.
If the Waveform to Search Is Set to a Logic Channel
• Waveforms of logic input modules
• Waveforms whose sub channel Value Type on the CAN bus monitor, CAN/CAN FD monitor, or
CAN & LIN bus monitor is Logic
• Waveforms whose sub channel on the SENT monitor is S&C or Error Trigger
Set the search source waveform
(a logic channel).
Set the polarity of each bit ( , , , - (don’t
use as a search condition)).
Set the number of times to
detect the conditions.
8.1 Searching for Edges