Teledyne ML – T101 Operation Manual
Operating Instructions
91
V
+DC Gnd
V OUT +
V OUT -
V IN +
V IN -
Recording
Device
ANALYZER
See Table 3-1 for
pin assignments
of Analog Out
connector on the
rear panel
Figure 4-4. Setup for Calibrating Analog Outputs
To make these adjustments, the
AOUT
auto-calibration feature must be turned off
(Section6.9). Activate the ANALOG I/O CONFIGURATION MENU (see Section
4.6.3), then press:
Press
SET>
to select the analog output channel to be configured:
DISPLAYED AS =
CHANNEL
CONC_OUT_1 =
A1
CONC_OUT_2 =
A2
TEST OUTPUT =
A4
DIAG
ANALOG I / O CONFIGURATION
PREV
NEXT
ENTR
EXIT
DIAG AIO
AOUTS CALIBRATED:
NO
< SET
SET>
CAL
EXIT
FROM ANALOG I/O CONFIGURATION MENU
DIAG AIO
CONC_OUT_1 :5V, NO CAL
< SET SET>
EDIT
EXIT
DIAG AIO
CONC_OUT_1 CALIBRATED: YES
< SET
CAL
EXIT
DIAG AIO
CONC_OUT_1 RANGE: 5V
SET>
EDIT
EXIT
DIAG AIO
CONC_OUT_2 CALIBRATED: NO
< SET
CAL
EXIT
DIAG AIO
CONC_OUT_1 VOLT–Z : 0 mV
U100 UP10 UP DOWN DN10 D100
ENTR EXIT
Toggle to increase / decrease the analog output
by 100, 10 or 1 counts.
Continue adjustments until the voltage measured
at the output of the analyzer and/or the input of
the recording device matches the value in the
upper right hand corner of the display to the
tolerance specified.
The concentration display
will not change. Only
the voltage reading of your voltmeter will change.
DIAG AIO
CONC_OUT_1 VOLT–S : 4500 mV
U100 UP10 UP DOWN DN10 D100
ENTR EXIT
DIAG AIO
CONC_OUT_1 REC OFS: 0 mV
< SET
SET>
EDIT
EXIT
If AutoCal is ON, go to
Section 6.7.3
EXIT
ignores the
new setting.
ENTR
accepts the
new setting.
DIAG AIO
CONC_OUT_1 AUTO CAL: OFF
< SET
SET>
EDIT
EXIT
Summary of Contents for T101
Page 2: ......
Page 3: ......
Page 4: ......
Page 6: ...ii This page intentionally left blank ...
Page 10: ...vi This page intentionally left blank ...
Page 12: ...viii This page intentionally left blank ...
Page 22: ...TABLE OF CONTENTS Teledyne ML T101 Operation Manual xviii This page intentionally left blank ...
Page 166: ...Remote Operation Teledyne ML T101 Operation Manual 162 This page intentionally left blank ...
Page 248: ......
Page 336: ...This page intentionally left blank B 4 07266B DCN6485 ...
Page 339: ...APPENDIX D Wire List and Electronic Schematics D 1 07266B DCN6485 ...
Page 340: ...This page intentionally left blank 07266B DCN6485 D 2 ...
Page 345: ...D 7 07266B DCN6485 ...