Teledyne API T/E-Series Models 101, 102, and 108 Software Documentation (PN05492D DCN6485)
A-15
Setup Variable
Numeric
Units
Default
Value
Value Range
Description
PHYS_RANGE1
PPM
2
0.1–2500
Low pre-amp range.
PHYS_RANGE2
PPM
20
0.1–2500
High pre-amp range.
CONC_RANGE1
Conc
500
0.1–50000
D/A concentration range 1.
CONC_RANGE2
Conc
500
0.1–50000
D/A concentration range 2.
CONC_RANGE3
6
Conc
500
0.1–50000
D/A concentration range 3.
150
OXY_FLOW_SET
4
cc/m
Warnings:
50–300
0–500
Nominal oxygenator flow set
point and warning limits.
OXY_FLOW_SLOPE
4
—
1
0.5–1.5
Slope term to correct
oxygenator flow rate.
700
SAMP_FLOW_SET cc/m
Warnings:
350–1200
0–1200
Sample flow set point for
flow calculation and warning
limits.
SAMP_FLOW_SLOPE
—
1
0.5–1.5
Sample flow slope correction
factor (adjusted flow =
measured flow x slope).
SFLOW_FILT_SIZE Samples
50 1–200
Sample
flow
adaptive
moving average filter size.
29.92
SAMP_PRESS_SET "Hg
Warnings:
15–35
0–100
Sample pressure set point
for pressure compensation
and warning limits.
CONV_TYPE
2
—
MOLY
NONE, MOLY Converter type.
315
CONV_SET
2
ºC
Warnings:
310–320
0–350
Converter temperature set
point and warning limits.
30
BOX_SET ºC
Warnings:
8–50
5–60
Box temperature warning
limits. Set point is not used.
7,
15
9
PMT_SET ºC
Warnings:
2–12,
2–20
9
0–40
PMT temperature set point
and warning limits.
07266B DCN6485
Summary of Contents for T101
Page 2: ......
Page 3: ......
Page 4: ......
Page 6: ...ii This page intentionally left blank ...
Page 10: ...vi This page intentionally left blank ...
Page 12: ...viii This page intentionally left blank ...
Page 22: ...TABLE OF CONTENTS Teledyne ML T101 Operation Manual xviii This page intentionally left blank ...
Page 166: ...Remote Operation Teledyne ML T101 Operation Manual 162 This page intentionally left blank ...
Page 248: ......
Page 336: ...This page intentionally left blank B 4 07266B DCN6485 ...
Page 339: ...APPENDIX D Wire List and Electronic Schematics D 1 07266B DCN6485 ...
Page 340: ...This page intentionally left blank 07266B DCN6485 D 2 ...
Page 345: ...D 7 07266B DCN6485 ...