Instrument Maintenance
Teledyne ML - T101 Operation Manual
210
Figure 8-4. Critical Flow Orifice Assembly
5.
Take out the components of the assembly: a spring, a sintered filter, two
O-rings and the critical flow orifice.
You may need to use a scribe or pressure from the vacuum port to get the parts out
of the manifold.
6.
Discard the two O-rings and the sintered filter.
7.
Replace the critical flow orifice.
8.
Re-assemble the parts as shown in Figure 8-4 using a new filter and O-
rings.
9.
Reinstall the NPT fitting and connect all tubing.
10.
Power up the analyzer and allow it to warm up for 60 minutes.
11.
Perform a leak check (refer to Section 9.5).